Sign in →

Test Code CHF8P Chromogenic Factor VIII Inhibitor Bethesda Profile, Plasma

Additional Codes

Epic# -

LAB000 (enter test name into comments)

 

Cerner name(s):

  • Miscellaneous Sendout


Ordering Guidance


This test is indicated for testing for FVIII inhibitors in patients being treated with the specific antibody emicizumab (Hemlibra).

 

This test is for detection of presence of specific inhibitors against factor VIII (FVIII). If the presence or type of inhibitor is unknown, APROL / Prolonged Clot Time Profile, Plasma or ALUPP / Lupus Anticoagulant Profile, Plasma should be ordered first.

 

Multiple coagulation profile tests are available. For testing that is performed with each profile, see Coagulation Profile Comparison.



Shipping Instructions


Send all vials in the same shipping container.



Specimen Required


Specimen Type: Platelet-poor plasma

Collection Container/Tube: Light-blue top (3.2% sodium citrate)

Submission Container/Tube: Plastic vials

Specimen Volume: 2 mL in 2 plastic vials, each containing 1 mL

Collection Instructions:

1. Specimen must be collected prior to factor replacement therapy.

2. If collecting sample through a port/line, be sure to waste the appropriate amount prior to collection.

3. For complete instructions, see Coagulation Guidelines for Specimen Handling and Processing.

4. Centrifuge, transfer all plasma into a plastic vial, and centrifuge plasma again.

5. Aliquot plasma (1 mL per aliquot) into 2 separate plastic vials leaving 0.25 mL in the bottom of centrifuged vial.

6. Freeze plasma immediately (no longer than 4 hours after collection) at -20° C or ideally, at or below -40° C.

Additional Information:

1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.

2. Each coagulation assay requested should have its own vial.


Forms

1. Coagulation Patient Information (T675)

2. If not ordering electronically, complete, print, and send an Coagulation Test Request (T753) with the specimen.

Secondary ID

610420

Useful For

Detecting the presence and titer of a specific factor inhibitor directed against coagulation factor VIII for patients on emicizumab (Hemlibra)

 

Detecting the presence and titer of an inhibitor directed against factor VIII

 

This test is not useful for detecting the presence of inhibitors directed against other clotting factors and will not detect the presence of lupus anticoagulants.

Profile Information

Test ID Reporting Name Available Separately Always Performed
CH8BI Chromogenic FVIII Inhibitor Interp No Yes
CHF8 Chromogenic FVIII, P Yes Yes
CH8B Chromogenic FVIII Inhibitor Titer,P No Yes

Method Name

Chromogenic

Reporting Name

Chromogenic FVIII Inhibitor Profile

Specimen Type

Plasma Na Cit

Specimen Minimum Volume

See Specimen Required

Specimen Stability Information

Specimen Type Temperature Time Special Container
Plasma Na Cit Frozen 14 days

Reject Due To

Gross hemolysis Reject
Gross lipemia Reject
Gross icterus Reject
IV heparin contamination Reject

Clinical Information

Factor VIII (FVIII) inhibitors are IgG antibodies directed against coagulation FVIII that typically result in development of potentially life-threatening hemorrhage. These antibodies may be alloimmune: developing in patients with congenital FVIII deficiency (hemophilia A) in response to therapeutic infusions of factor VIII concentrate or autoimmune: occurring in patients without hemophilia (not previously factor VIII deficient) either spontaneously, during pregnancy, or in association with autoimmune diseases.

Reference Values

CHROMOGENIC Factor VIII Activity Assay

Adults: 55.0-200.0%

Normal, full-term newborn infants or healthy premature infants usually have normal or elevated factor VIII.*

*See Pediatric Hemostasis References section in Coagulation Guidelines for Specimen Handling and Processing.

 

BETHESDA TITER

≤0.5 Bethesda Units

Interpretation

An interpretive report will be provided when testing is completed, noting a presence or absence of a chromogenic factor VIII inhibitor.

Cautions

Contamination with excess heparin and hemodilution due to improper specimen collection through an intravenous access device or collection above a running intravenous fluid line may cause spurious results.

Clinical Reference

1. Peyvandi F, Oldenburg J, Friedman KD. A critical appraisal of one-stage and chromogenic assays of factor VIII activity. J Thromb Haemost. 2016;14(2):248-261

2. Verbruggen B, van Heerde WL, Laros-van Gorkom BA. Improvements in factor VIII inhibitor detection: From Bethesda to Nijmegen. Semin Thromb Hemost. 2009;35(8):752-759

3. Miller CH, Platt SJ, Rice AS, Kelly F, Soucie JM: Hemophilia Inhibitor Research Study Investigators. Validation of Nijmegen-Bethesda assay modifications to allow inhibitor measurement during replacement therapy and facilitate inhibitor surveillance. J Thromb Haemost. 2012;10(6):1055-1061

Method Description

The Chromogenic Factor VIII assay is performed on the Instrumentation Laboratory ACL TOP 700 using the CRYOcheck Chromogenic Factor VIII kit. In this 2-stage assay, patient plasma is diluted and combined with reagents containing bovine factor X, human factors IXa and IIa, calcium chloride, and phospholipids. The factor VIII in the patient's plasma aids in the activation of factor X to factor Xa. After a specified incubation period, chromogenic substrate is added at which time, the factor Xa, present from the previous step, hydrolyzes the substrate into peptide and p-nitroaniline (pNA). The color produced by the release of pNA is measured photometrically at 405 nm and is proportional to the factor VIII in the sample.(Package insert: CRYOcheck Chromogenic Factor VIII. Precision BioLogic Inc; Rev V04, 06/2020)

 

In the Bethesda procedure, patient plasma is heat inactivated (HI) at 56° C for 30 minutes. Next using the HI patient plasma, serial dilutions are prepared and mixed in equal volumes with normal pooled plasma. The mixture is incubated 2 hours at 37° C. At the end of the incubation, chromogenic factor VIII (CHF8) activity is measured and compared to a control performed at the same time. The difference between the CHF8 activity of the patient's incubation mixture and that of the control is used to calculate the titer. The residual CHF8 activity is converted to Bethesda units: 50% residual CHF8 is equal to 1 Bethesda unit. Assays using the same basic principle as the Bethesda assay are used to quantitate the inhibitors of the other coagulation factors.(Kasper CK, Aldedort LM, Counts RB, et al. A more uniform measurement of factor VIII inhibitors. Thromb Diath Haemorrh. 1975;34[03]:869-872. doi:10.1055/s-0038-1651378; Miller CH. Laboratory testing for factor VIII and IX inhibitors in haemophilia: A review. Haemophilia. 2018;24[2]:186-197. doi:10.1111/hae.1342)

Day(s) Performed

Monday through Friday

Report Available

3 to 4 days

Specimen Retention Time

7 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

See Individual Test IDs

CPT Code Information

CHF8-85130

CH8B-85335

CH8BI-85390-26

LOINC Code Information

Test ID Test Order Name Order LOINC Value
CHF8P Chromogenic FVIII Inhibitor Profile 95121-0

 

Result ID Test Result Name Result LOINC Value
CH8B Chromogenic FVIII Inhibitor Titer,P 93450-5
606844 Chromogenic FVIII Inhibitor Interp 95122-8
CHF8 Chromogenic FVIII, P 49865-9
606865 Reviewed by 18771-6